A-A-50734B
TABLE I. DC Voltage.
* Least Significant Digit.
2.6.2.2 Input Impedance. The input impedance shall be greater than 9.9 Megohms.
2.6.2.3 Overload. The multimeter shall be capable of continuously withstanding, without damage or blown fuses, 1,000 Vrms on all ranges, except 500 Vrms continuously on the 300.0 mV range.
2.6.2.4 Normal mode rejection ratio. Normal mode noise rejection shall not be less than 60 decibels (dB) at 50 and 60 Hz (±0.15%).
2.6.2.5 Common mode rejection ratio. Common mode rejection, with a 1,000 Ohms unbalanced in the low lead, shall not be less than 120 decibels (dB) at 50 and 60 Hz (±0.15%).
2.6.3 AC Voltage.
2.6.3.1 Range, resolution, and accuracy. Minimum performance shall be as specified in Table
II.
TABLE II - AC Voltage.
RANGE |
RESOLUTION |
ACCURACY |
||
320.0 mV 3.200 V 32.00 V 320.0 V 1000 V |
100 µV 1 mV 10 mV 100 mV 1V |
40 Hz to 2 KHz |
2 KHz to 10 KHz |
10 KHz to 30 KHz |
±(0.5%+3 LSD) ±(0.5%+3 LSD) ±(0.5%+3 LSD) ±(0.5%+3 LSD) ±(1% + 3 LSD) |
±(2%+3 LSD) ±(2%+3 LSD) ±(2%+3 LSD) ±(2%+3 LSD) ±(3%+3 LSD) |
±(4%+10 LSD) ±(4%+10 LSD) ±(4%+10 LSD) ±(4%+10 LSD) Not to exceed 15% |
2.6.3.2 Input impedance. The input impedance shall be greater than 9.9 Megohms.