A-A-53007
Table II. Tests (Continued).
Linearity: Bipolar triangle, 2 Hz, 4 mVpp ECG rate meter/alarm test: 60-30-60-120-240-60 Hz automatic cycle sequence, 20 second cycle duration each. |
Examination. Each defibrillator analyzer shall be examined to determined compliance with the classification of defects in table III.
Table III. Classification of Defects.
Categories |
Defects |
Major |
|
101 |
Power and control switch (s) non-functional |
102 |
Digital indicator non-functional |
103 |
Inputs and outputs non-functional |
Minor |
|
201 |
Identification markings not complete, permanent, and legible. |
202 |
Finish not free of chips, cracks, scratches and other appearance defects. |
203 |
Any component part not secured in place. |
Examination shall not be limited to the classified possible defects listed herein.
Contractor certification. The contractor shall certify that the product offered meets the salient characteristics of this description and conforms to the producers’ own drawings, specifications, standards, and quality assurance practices. The Government reserves the right to require proof of such conformance prior to first delivery and thereafter as may be otherwise provided for under the provisions of the contract.