INCH-POUND A-A-50740A
8 July 2009
SUPERSEDING A-A-50740
30 October 1987
COMMERICIAL ITEM DESCRIPTION SPECTRUM ANALYZER
(10 kHz - 50 GHz)
Reinstated after (8 July 2009) and may be used for new and existing designs and acquisitions.
The General Services Administration has authorized the use of this commercial item description by all federal agencies.
1. SCOPE. This Commercial Item Description(CID) describes a solid-state, internal pre- selection, programmable, 10 Hz resolution bandwidth, 10 kHz to 50 GHz, Spectrum Analyzer including external mixers. This CID is meant as a minimum requirement for the Spectrum Analyzer in which only those manufacturers that meet or surpass the following requirements are qualified per this CID. Test equipment offered may be commercially available and may be modified to the extent necessary to meet the following requirements are supplied per this CID.
2. CLASSIFICATION. Equipment covered by this CID shall be commercially available equipment and may be modified to the extent necessary to meet the following description. The equipment shall be Class 3, in accordance with MIL-PRF-28800F, except as specified herein.
3. SALIENT CHARACTERISTICS. The equipment shall be capable of operation within the accuracies, limits, and specifications herein.
3.1 Safety and Environmental. The equipment shall meet all safety and environmental requirements as specified in MIL-PRF-28800F for the classification as stated herein except as specified herein.
3.2 Temperature Requirements. The temperature ranges for both operating and non- operating conditions shall be as specified for Class 3 equipment in MIL-PRF-28800F.
3.3 Electrical Power Sources. The equipment shall operate from nominal commercial, military, and shipboard power sources of 110/220 Volts AC (+10%) at line frequencies of 45 Hz to 440 Hz. Power consumption shall not exceed 650 VA. Transient state conditions shall meet MIL-PRF-
28800F requirements for Class III equipment.
3.4 Magnetic Environment. The equipment shall comply with the magnetic environment requirements of MIL-PRF-28800F.
3.5 Reliability. The design of the equipment shall be such that under normal use and operation the equipment does not fail within 8000 hours of operation with a statistical certainty of 95%.
Beneficial comments, recommendations, additions, deletions, clarifications, etc. and any other data that may improve this document should be sent to: WR-ALC/742 CBSG, 460 Richard Ray BLVD, Ste 200, Robins AFB, GA 31098-1611. Since contact information can change, you may want to verify the currency of this address information using the ASSIST Online database at https://assist.daps.dla.mil .